INTEGRA 3 MICROMANIPULATION SYSTEM WITH 2XSAS-SE Air Syringes with equilibration button 1XINTEGRA 3 microscope stage

Category:

Product Specifications

Heating System Thermosafe™ Air Heating System with Heated Metal Insert:
25mm aperture (standard)
16mm aperture (optional)
ITO Heated Glass Insert (optional)
Temperature Controller Accuracy: better than ±0.1ºC when calibrated against known reference
Resolution: 0.1ºC
Setpoint Temperature Range: 30-45ºC
Display 3.5″ colour LCD touch screen display for temperature control
Connectivity USB socket for connection to PC running RI Viewer™ software
Syringes Choice of SAS (air), SAS-SE (air) and SOS (oil)
Weight Maximum weight of microscope mounted components (excluding front and rear adaptors): 10.4kg
Actual weight is configuration dependent
Dimensions Footprint (not including microscope): width 56cm, depth 38cm
Power Supply Input: 100-240VAC, 50-60HZ, Max. 1.8A, Class 1
Output: 12VDC, Min. 11A (132W)
Operating Range Temperature: 15ºC (59ºF) to 40ºC (104ºF)
Humidity:15% to 85% RH (non condensing)
Micropipette Compatibility Compatible with CooperSurgical micropipettes (or other industry standard 1.0mm outside diameter micropipettes or equivalent specification).